Beamline Operations


Safety in Experimental Hutches

Certain safety mechanisms and restrictions are in place to reduce the likelihood of accidents while working in or around the experimental hutches.

Experimental Hutch Doors

Doors to experimental hutches must be properly secured before an experiment can be conducted.

  • Beamline 08B1-1 has two swinging hutch doors. The first (on the left when looking at the hutch from outside) must be closed and the latch locked before closing the second door (on the right when looking at the hutch from outside). As the doors are very heavy, care must  be taken to close them slowly and gently. To open the door, press the Safety Shutters "Close" button on the panel labelled "SOE-1", followed by the Main Door "Unlock" button (see diagram on left below).
  • Beamline 08ID-1 has a main door to the hutch that is controlled electronically. A panel labelled "SOE-1" beside the door contains Main Door "open" and "close" buttons (see diagram on right below). In case of emergency the door can be stopped by pressing the large red "Door Stop" button. To open the door, press the Shutters "Close" button, then hold down the Main Door "Open" button.

 

  • If you ever find yourself locked inside a hutch with the doors closed, you can disable the beam by pushing the red button on the panel labelled "Emergency Beam Shut-Off" inside the hutch immediately next to the doors. This prevents any possibility of beam entering the experimental hutch. 
  • Doors can be opened from inside by pressing a green "Push To Exit" button.

The control panels outside the hutches also contain switches that the Floor Coordinator or Health, Safety and Environment can use to disable the beamlines.

  • If either of these are disabled or show a major fault, contact the Floor Coordinator or beamline staff.

Detector Safety

  • For your safety, detectors should be moved before entering a hutch in order to avoid accidents.
  • It is important not to touch the surface of detectors or leave any objects in the path of a detector or on the platform which could potentially damage a detector.
  • Do not adjust 2-theta. Contact staff if you require 2-theta to be moved.
  • Minimum detector distances:
    • Beamline 08B1-1: 115 mm
    • Beamline 08ID-1: 150 mm

Automounter Safety

  • The automounters on both CMCF beamlines are available to users. Please contact staff prior to your arival if you wish to use them and read about Samples & Automounters.
  • When an automounter is in operation, a caution sign will be posted on the hutch door and an orange light may be visible on top of the automounter. During these times only those specifically authorized may enter the hutch. The arm of the automounter may make unexpected movements when in operation.
  • Do not mount samples using the automounter while anyone is inside the hutch.
  • Please allow staff to load your cassettes/Uni-Pucks into the automounter and probe them for you.

 

Mounting Samples

The endstations have various sensitive components that surround the sample positions, as depicted below.

  • To mount a sample manually, simply click "Mounting Mode" in the Beamline Setup tab of MxDC before entering the hutch. This moves the detector, capillary, beamstop, backlight and cryojet away from the sample position to give you enough room to safely mount the sample without accidentally hitting any components. Although these components move away, they could still be easily damaged if care is not taken during mounting. Be particularly observant of the capillary on each beamline.
  • Samples should only be mounted or dismounted in Mounting Mode.
  • On beamline 08B1-1, various components such as the backlinght and capillary are retracted and protected by covers. These covers should not be obstructed. The capillary should not be visible during mounting.
  • Stay clear of the sample position and detector when switching to "Centering Mode".
  • Cryo-tools and other mounting tools are available near the beamlines and in the Sample Preparation Room.
  • Safety gloves, glasses, nitrogen dewars and microscopes are available in the 08ID-1 Control Area and the Sample Preparation Room (see map).

Hutch Lockup

Before locking up any Experimental Hutch, it is important to ensure that no objects are left resting on or near the beamline.

  • Objects near the base of the detector or near the sample area may potentially interfere with movements of motors.

Enclosure Search

Before locking up a hutch, it must be searched to ensure there is no one inside.

  • There is one green Enclosure Search button in the Beamline 08B1-1 experimental hutch and 2 in the 08ID-1 experimental hutch, each located on the bottom of grey boxes. 
  • For Beamline 08B1-1, push this button while sweeping through the hutch to ensure no one is inside.
  • For Beamline 08ID-1, push the button nearest the endstation first, sweep through the hutch and press the second one located in the rear corner, ensuring there is no one else in the hutch or behind any equipment.

  • Once the hutch has been searched and enclosure search button(s) pressed, exit and close the hutch door.
    • Beamline 08B1-1: gently close the door, you should hear the buzzers sound inside.
    • Beamline 08ID-1: On the SOE-1 Panel, located immediately beside the door on the exterior of the hutch, press and hold the Main Door "close" button until you can hear the buzzers sound inside.
  • See the section on hutch doors for additional information.


Preparing for Data Collection

Loggin In, Opening MxDC, Directory Structure

Log into one of the user terminals and open MxDC. If you are collecting remotely, you will receive instructions by e-mail ahead of time for the procedure.

When you open a new session of MxDC, a new directory with the date will be created. We recommend that you use a single directory like this for your data collection run, under which all your individual data sets will be collected in sub-directories. This facilitates data transfer to a hard drive using the archiver script. MxDC will assume that you will use the directory that it created for a week. After a week has elapsed, upon opening MxDC, the system will create another new directory with a new date.

See the MxDC (Data Collection) page for detailed information about using MxDC.

Data Collection using Automounter

  1. You will be contacted by your local contact prior to beginning a session with the automounter.
  2. For the first mount, ensure that there is no sample on the goniometer before starting. You can visualize the sample position using the Hutch Camera video feed located on the Beamline Setup tab. 
  3. From the Samples tab, mount your chosen sample (see Samples tab section of MxDC for more information)
  4. After the automounter mounts your sample, the beamline should automatically go into Centering Mode. Occasionally this may fail, in which case wait until the automounter has completed mounting your sample, then click the Centering Mode button.
  5. Align the crystal using the "Samples" tab of MxDC. It is important that you rotate the crystal to ensure it is centered in all directions.
  6. You may need to click "Restore Beam" on the "Beamline Setup" tab of MxDC to open the shutters if there has been an injection or the hutch has just been secured.
  7. In the "Data Collection" tab of MxDC, enter parameters such as the directory you wish to store your data in, prefix for the image file names, distance of detector, omega delta, start frame and number of frames or degrees.
  8. Click "Save" before collecting.  The endstation will automatically switch to "Collect Mode" (the backlight will move out of the way) before the first image is collected.
  9. To change samples, simply tell the automounter to mount the next one. The system will automatically go into "Mounting Mode" and dismount the sample that it mounted previously before mounting the chosen sample.
  10. Once you are finished your last sample, dismount it, close MxDC and inform staff by e-mail that you have finished.

Manual Data Collection Procedure

  1. From the "Beamline Setup" tab of MxDC, click "Mounting Mode". You must be in Mounting Mode to mount or change samples manually in order to avoid damaging the equipment. If you need more room, please move the detector to a greater distance prior to mounting (400 - 500 mm is ideal).
  2. Carefully mount the sample onto the goniometer (see Mounting Samples).
  3. From the "Beamline Setup" tab of MxDC, click "Centering Mode".
  4. Align the crystal using the "Samples" tab of MxDC. It is important that you rotate the crystal to ensure it is centered in all directions.
  5. Lock-up the endstation (see Hutch Lockup).
  6. Click "Restore Beam" on the "Beamline Setup" tab of MxDC to open the shutters.
  7. In the "Data Collection" tab of MxDC, enter parameters such as the directory you wish to store your data in, prefix for the image file names, distance of detector, omega delta, start frame and number of frames or degrees.
  8. Click "Save" before collecting.  The endstation will automatically switch to "Collect Mode" (the backlight will move out of the way) before the first image is collected.
  9. Before changing samples, click "Mounting Mode".
  10. Once you are finished your last sample, dismount it, close MxDC and inform staff by e-mail that you have finished.

Energy Changes

  • 08B1-1: energy changes are fully automated between 6 - 18 keV
  • 08ID-1: energy changes partially automated between about 6 - 18 keV. Carefully ensure that the beam is optimized using Beamtuner after an energy change (08ID-1 only)
  • if you change the energy, allow about 1-2 minutes for the beamline optical components to adjust (or before beginning manual optimization using Beamtuner on beamline 08ID-1)
  • If you require energies outside the ranges outlined above, please contact staff for help during regular hours.

Beam Optimization

After an injection or after locking up the hutch, click Restore Beam from the Beamline Setup tab of MxDC. You may also need to do this at the beginning of your remote session.

  • 08B1-1: the beam is kept optimized automatically on beamline 08B1-1
  • 08ID-1: use Beamtuner to ensure the beam remains optimized on beamline 08ID-1

Beamtuner (beamline 08ID-1 only). Careful optimization must be ensured using Beamtuner on beamline 08ID-1. To optimize the beam, open a local terminal and type "beamtuner" to open the Beamtuner window (leave it open during your collection for easy access). Wait until any energy change is completed and the red line is somewhat stable before making adjustments (as in figure, right). Use the "+++/---" buttons to make manual adjustments until the observed value and red readout line are at peak values.  

Occasionally, the reading may become erratic  and nonsensical. This can be solved by clicking the Reset button.

NOTE: Do not make manual adjustments if an energy change is in progress, and avoid it during your data collection. The optimization should be checked every couple of hours.

Occasionally you may still not have full beam after an injection is finished, even after completing the above instructions. Check the Beamline Setup Tab carefully to ensure all Status Checks are green and there are no issues. Also ensure you have clicked "Restore Beam". If you are still not getting full beam and there are no error warnings, try changing the energy slightly in the Beamline Setup Tab (for example if you were using 12.658 keV, try changing it slightly to 12.650 keV). This normally corrects the issue unless something more serious is wrong.